Compendium: Technical specification for high frequency electrosurgical generator ARC 300 |
Max. output power for monopolar Cut |
300 W |
Max. output power for monopolar Coag |
120 W |
Max. output power for bipolar Coag |
120 W |
NE safety monitoring (EASY) |
Yes |
Rated frequency |
330 kHz sinusoidal / 1 MHz pulse |
Mains connection / max. current consumption |
90 V - 110 V / 6.4 A |
198 V - 260 V / 3.2 A |
Mains fuse |
2x 90 V - 110 V / 8 A slow-acting |
2x 198 V - 260 V / 4 A slow-acting |
Mains frequency |
50 Hz / 60 Hz |
Power consumption in standby mode |
25 watts |
Max. power consumption at max HF power |
500 watts |
Equipotential bonding connector |
Yes |
Storage and transport conditions |
Temperature |
-20 °C - +50 °C |
Relative humidity |
0 - 75 %, non-condensing |
Air pressure |
500 - 1600 mbar |
Operating conditions |
Temperature |
10 °C -+40 °C |
Relative humidity |
30 - 85 %, non-condensing |
Air pressure |
700 - 1600 mbar |
Protection class to |
EN 60601-1 I |
Type according to |
EN 60601-1 CF |
Operating mode |
10 sec / 30 sec |
|
Classification according to EC Directive 93/42/EEC |
Class IIb |
Dimensions and weights |
External dimensions (W x H x D) |
430 x 180 x 400 mm |
Weight |
14 kg |